WSA-P200A [Specification / Optional]

Specification
Content Description
Display Range 1mTorr~20Torr
Error Range15%
DisplayFND, LED
OperationFront & button type
CalibrationUser can calibrate by using SPAN & ZERO dial
Set PointDual Set Point (High/Low Set Value) Relay Contant
Set PointDual Set Point (High/Low Set Value) Relay Contant
Date Gathering Cycle200ms
Digital Output12bit output of current vacuum level
Monitoring ProgramRS232C data transfer to PC program
ADC12Bit
Sensor Type15mm O.D.(standard)
Sensor Cable2M(default length)
Power20W. AC85~260V,60/50Hz


Specification
Content Description
Sensor Cable5m, 10m
Fitting18mm O.D., ISO KF(NW16, NW25), CF
Gauge Adaptor(T)NW25~15 phi Gauge port-NW25 / Venting
Gauge Adaptor(1)NW25~15 phi Gauge port / Venting


WSA-P200A[Function]

  • Front Panel

    ¨ç Digital Display : LED display by 0.1 point
    ¨è Unit Change Key : Torr / Pascal
    ¨é Set Point : Setting 2 set points for auto controlling
    -Set Point can be changed by using the following key : RNG, Up, Down, ENT
    ¨ê MODE : User defined type mode setting
    ¨ë Filament On / Off : ON Key

  • Rear Panel

    ¨ç PC monitoring by using RS232C
    - Monitoring Graph can be saved by BMP file
    - Monitoring Data can be saved MS Excel format
    ¨è Digital Output Terminal : Real time output with 12bit data
    ¨é Relay Terminal
    Set Point Relay Output(SPDT) : 2 Point
    ¨ê User Calibration
    - SPAN : Calibration within 19~17 torr
    - ZERO : Display LO at 9x10-4torr
    ¨ë Power : Free Volts(85~260V, 60/50Hz), On/Off Switch, and Fuse for safty

WSA-P200A[Setting]

key Description
¡¤ ON : filament on/off.
Starting display by press[ON] button
¡¤ UNIT : Changing Display Unit between Torr / Pa
Initially Torr is default unit
¡¤ MODE : Optional function key for user setting
¡¤ SET : Setting the set point(S.1 S.2 moving)
¡¤ RNG : Moving the digit Range
¡¤ ENT : Saving current setting
¡¤ UP/DOWN : Set point up and down


Set Point
- Set Value Range : 19~0.001 torr(2500~0.133 Pa)
- By pressing SET button you can move SET1 modr, SET2 mode and Current Vacuum mode.
S.1 LED display means SET1 mode and S.2 LED display means SET2 mode.
-[RNG] button can move the digit postion(only works at SET mode)
- By using [UP]/[DOWN] button, you can change the value for setting your favorite SET point
- By Pressing [ENT] button you can save the SET Value.


Dimension & Sensor Fittings

1. Controller







¢Â Supported Fittings ¥°(without Adaptor)
¡á 15mm O.D. (Standard)
¢Â Supported Fittings ¥±(by Adaptor)
¡á 18mm O.D. (Standard)
¡á KF(NW) 16
¡á CF Type(1.33 inch)

¢Â Supported Fittings ¥²
¡á Gauge Adaptor(T) : NW25-15¢± Gauge Port-NW25
¡á Gauge Adaptor(1) : NW25-15¢± Gauge Port
2. Sensor


Assignment

Accessory Terminal


PIN NO signal PIN NO signal
1 GND 14 Sign(+/-)
2 Digital output A1(20) 15 Not Use
3 Digital output A2(21) 16 Not Use
4 Digital output A3(22) 17 Set 1 NO
5 Digital output A4(23) 18 Set 1 COM
6 Digital output B1(20) 19 Set 1 NC
7 Digital output B2(21) 20 Set 2 NC
8 Digital output B3(22) 21 Set 2 COM
9 Digital output B4(23) 22 Set 2 NO
10 Digital output C1(20) 23 Not Use
11 Digital output C2(21) 24 GND
12 Digital output C3(22) 25 GND
13 Digital output C4(23)


Sensor Cable Pin Connection


WSA(WooSung Automa Co. Ltd.,)

  • Information

    Address : 646 16, Seokpo 5 Ri, Jangan Myeon, Hwaseong Si, Kyoungkji Do, Korea
    TEL : +82 31 358 9640
    FAX : +82 31 358 9644
    URL : http://www.wsautoma.co.kr

  • Other product Line

    2 stage oil rotary pump
    W2V Series : 100, 200, 400, 600, 800(L/min)
    SV Series : SV-145
    Diffusion pump, Booster pump, Valve, .....
    Vacuum Accessory
    O-ring, Oil Mist Trap, Bellows, .... etc.



INFORMATION
  • »óÈ£¸í : ¿ì¼ºÁø°ø»ó»ç
  • ´ëÇ¥ÀÚ¸í : ±è±âÈì
  • ÁÖ¼Ò : °æ±âµµ ±ºÆ÷½Ã »êº»·Î77¹ø±æ 44
  • ´ëÇ¥ÀüÈ­ : 031-427-0936, 031-424-0933
  • ´ëÇ¥ÀüÈ­ : 031-427-0936, 031-424-0933
  • Æѽº: 031-427-7043
  • Æѽº: 031-427-7043
  • E-mail : wsvp2002@hanmail.net
  • E-mail : wsvp2002@hanmail.net
  • ´ëÇ¥ÀüÈ­ : 031-427-0936, 031-424-0933
»ç¾÷ÀÚµî·Ï¹øÈ£
  • »ç¾÷ÀÚµî·Ï¹øÈ£ : 138-03-28338
°³ÀÎÁ¤º¸º¸È£Á¤Ã¥ Ã¥ÀÓÀÚ
  • e-mail : wsvp2002@hanmail.net
  • °³ÀÎÁ¤º¸º¸È£Á¤Ã¥ Ã¥ÀÓÀÚ : ±è±âÈì
SITEMAP
¸ðµç ÄÁÅÙÃ÷ÀÇ ¹«´Üº¹Á¦ ¹× ÀçÆǸŸ¦ ±ÝÁöÇÕ´Ï´Ù.
Copyright(c) 2021 All Rights Reserved. ¿ì¼ºÁø°ø»ó»ç